Implant-Prosthodontic Discrepancy of Complete-Arch Cobalt-Chromium Implant Frameworks Manufactured Through Selective Laser Melting Additive Manufacturing Technology Using a Coordinate Measuring Machine

Marta Revilla-Len, DDS, MSD/Laura Ceballos, DDS, PhD/Mutlu zcan, DDS, Dr Med Dent, PhD


PMID: 30892285
DOI: 10.11607/jomi.6739

Purpose: The objective was to measure the implant-prosthodontic discrepancy of complete-arch implant-supported frameworks made of cobalt-chromium (Co-Cr) fabricated using selective laser melting (SLM) additive manufacturing technologies. Materials and Methods: A completely edentulous maxillary cast with seven implant replicas was obtained. Co-Cr SLM frameworks (n = 9) from three different providers (SLM-1, SLM-2, SLM-3) were manufactured. A coordinate measuring machine was selected to measure the implant-prosthodontic discrepancy (μm) on the x-, y-, and z-axes and the 3D gap (3D = √x2 + y2 + z2 where implants were considered as the statistical unit (n = 7). One-way analysis of variance (ANOVA), Student-Newman-Keuls, and Tukey tests were used to analyze the data (α = .05). Results: The mean 3D implant-prosthodontic discrepancy (μm) was higher for SLM-1 (73.77 27.94) than for SLM-2 (47.54 22.63) and SLM-3 (47.26 22.57). At the x-axis, SLM-2 showed a significantly smaller gap (16.21 9.6) than SLM-3 (32.92 27.77) and SLM-1 (34.77 21.85). At the y-axis, however, SLM-3 presented a significantly smaller gap (27.97 9.49) than SLM-2 (38.84 27.82) and SLM-1 (54.35 29.89). Similarly, at the z-axis, SLM-3 (4.01 2.29) showed the least gap, followed by SLM-2 (9.09 7.63), which was significantly smaller than that of SLM-1 (16.14 21.09). Conclusion: The three SLM additive manufacturing technologies tested showed implant-prosthodontic discrepancies ranging from 4.01 to 54.35 μm, which could be considered in the clinically acceptable range. Distortion at the z-axis was significantly less compared with the x- and y-axes in all the groups tested.


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2020 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help