Accuracy Assessment of Image-Guided Implant Surgery: An Experimental Study

Jürgen Hoffmann, MD, DMD, PhD/Carsten Westendorff, MD/Michael Schneider/Siegmar Reinert, MD, DMD, PhD


PMID: 15973949

Purpose: To accurately accomplish the drilling of an implant socket, the use of image-guided navigation has become an option. The aim of this study was to evaluate the 3-dimensional (3D) accuracy of navigation-guided drilled holes. Materials and Methods: Laboratory accuracy measurements were obtained on an acrylic resin model with standardized target holes drilled by a computerized numerical control machine. The model was scanned by a multislice computerized tomography scanner and registered with fiducial marker–based algorithms. Navigated drillings were performed using an optical navigation system based on passive marker technology. Coordinates of drilled holes were determined by a 3D-digitizer probe, and accuracy was assessed for all 5 degrees of freedom using a computer-aided design system (Pro/Engineer). Results: A total of 240 drillings were evaluated. Mean registration error was 0.86 mm (SD 0.25 mm). Target point deviation between preplanned and actual drill starting point was 0.95 mm (SD 0.25 mm). The deviation in terms of full length was 0.97 mm (SD 0.34 mm), and mean angular deviation on the coronal and sagittal planes was 1.35 degrees (SD 0.42 degrees). Discussion: The accuracy of image-guided navigation depends on imaging modalities, patient-to-image registration procedures, and instrument tracking. The technical accuracy and the navigation procedure, as evaluated in the study presented, seem to be of minor influence. Conclusion: The data obtained by this in vitro study demonstrate that the accuracy of navigation-based drilling may be sufficient for clinical practice, particularly in terms of the transferability of preplanned trajectories. However, in vivo clinical trials need to be performed to evaluate the clinical accuracy and treatment quality of navigation-guided interventions. Int J Oral Maxillofac Implants 2005;20:382–386


Full Text PDF File | Order Article

 

 
Get Adobe Reader
Adobe Acrobat Reader is required to view PDF files. This is a free program available from the Adobe web site.
Follow the download directions on the Adobe web site to get your copy of Adobe Acrobat Reader.

 

© 2022 Quintessence Publishing Co, Inc JOMI Home
Current Issue
Ahead of Print
Archive
Author Guidelines
About
Accepted Manuscripts
Submission Form
Submit
Reprints
Permission
Advertising
Quintessence Home
Terms of Use
Privacy Policy
About Us
Contact Us
Help